Research Article | | Peer-Reviewed

Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy

Received: 13 September 2026     Accepted: 8 October 2026     Published: 28 October 2026
Views:       Downloads:
Abstract

Postoperative vital sign monitoring is essential for early detection of inpatient hemodynamic and respiratory complications. Although conventional bedside monitors serve as the clinical gold standard, their fixed design limits patient mobility and continuous ambulatory monitoring, which hinders early rehabilitation after orthopedic surgery. Wearable monitoring devices have shown great potential for postoperative vital sign surveillance due to their mobility and continuous data collection capabilities, yet real-world validation data in orthopedic postoperative patients remain insufficient, and wearable blood pressure measurement accuracy remains controversial. This study aimed to evaluate the agreement between a wearable monitor and a standard bedside monitor for heart rate, respiratory rate, systolic blood pressure, diastolic blood pressure, and peripheral oxygen saturation (SpO2) in patients undergoing knee arthroscopy. A total of 100 elective knee arthroscopy patients were prospectively enrolled. Paired vital sign data were collected simultaneously via the two devices at four daily time points over two postoperative days. Intra-class correlation coefficient (ICC), Lin’s concordance correlation coefficient, Bland-Altman analysis, temporal trend analysis, and clinical acceptability evaluation were performed for statistical assessment. After excluding five missing values, 3995 paired observations were analyzed. The wearable device showed excellent agreement for heart rate (ICC = 0.994, bias = −0.63 bpm, 95% LoA: −2.97 to 1.70 bpm), good agreement for respiratory rate (ICC = 0.940, bias = −0.51 breaths/min, 95% LoA: −2.71 to 1.68 breaths/min), and moderate-to-good agreement for SpO2 (ICC = 0.872, bias = −0.85%, 95% LoA: −3.50 to 1.80%). Diastolic blood pressure showed moderate agreement (ICC = 0.827, bias = −1.42 mmHg, 95% LoA: −12.90 to 10.06 mmHg), while systolic blood pressure demonstrated the lowest agreement (ICC = 0.811) with a prominent positive systematic bias of 5.63 mmHg. Temporal trends remained stable across postoperative days, with consistent inter-device bias over time. The clinical acceptability rates of heart rate, respiratory rate and SpO2 exceeded 99.5%, compared with 92.2% for diastolic blood pressure and only 79.6% for systolic blood pressure. In conclusion, the wearable monitor provides reliable heart rate, respiratory rate, and SpO2 measurements and is suitable for routine postoperative monitoring after knee arthroscopy. However, blood pressure measurements, particularly systolic blood pressure, show notable systematic deviation, and targeted algorithm calibration is needed to improve its blood pressure monitoring performance.

Published in Clinical Medicine Research (Volume 15, Issue 4)
DOI 10.11648/j.cmr.20261504.12
Page(s) 60-68
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2026. Published by Science Publishing Group

Keywords

Wearable Monitoring, Vital Signs, Bedside Monitor, Measurement Agreement, Knee Arthroscopy, Postoperative Monitoring

References
[1] Khanna AK, O'Connell NS, Saha AK, Hicks MH, Weller RS, Harris L, Cusson BD, Faris A, Huffman CS, Segal S, Wells BJ, Kirkendall ES, Sessler DI. Continuous vs Intermittent Postoperative Vital Sign Monitoring: A Cluster Randomized Crossover Trial. JAMA Netw Open. 2026 Mar 2; 9(3): e263290.
[2] Mølgaard J, Grønbæk KK, Rasmussen SS, Eiberg JP, Jørgensen LN, Achiam MP, Rohrsted M, Singh UM, Hoang TH, Søgaard M, Meyhoff CS, Aasvang EK. Continuous Vital Sign Monitoring at the Surgical Ward for Improved Outcomes After Major Noncardiac Surgery: A Randomized Clinical Trial. Anesth Analg. 2025 Oct 1; 141(4): 807-817.
[3] Passos RDH, Andari LVWBU, Assuncão MSC. Monitor smart, use better: the future of haemodynamic monitoring. J Clin Monit Comput. 2024 Dec; 38(6): 1437-1439.
[4] Dash S, Bandyopadhyay A, Puri S, Yalla B. The role of wearable devices in pre-anaesthesia check-up facilitation: A scoping review. J Perioper Pract. 2026 Oct; 36(10): 581-589.
[5] Lee J, Kong S, Shin S, Lee G, Kim HK, Shim YM, Cho J, Kang D, Park HY. Wearable Device-Based Intervention for Promoting Patient Physical Activity After Lung Cancer Surgery: A Nonrandomized Clinical Trial. JAMA Netw Open. 2024 Sep 3; 7(9): e2434180.
[6] El Arab RA, Al Moosa OA, Sagbakken M, Ghannam A, Abuadas FH, Somerville J, Al Mutair A. Integrative review of artificial intelligence applications in nursing: education, clinical practice, workload management, and professional perceptions. Front Public Health. 2025 Aug 1; 13: 1619378.
[7] Flick M, Bergholz A, Sadtler H, Kouz K, Hoppe P, Auinger J, Krause L, Zöllner C, Thomsen KK, Saugel B. Abnormal vital signs during recovery from noncardiac surgery on general wards: A blinded prospective observational study. Eur J Anaesthesiol. 2026 Sep 1; 43(9): 769-778.
[8] Rincon F, Pidoux J, Murali S, Goy JJ. Performance of the new SmartCardia wireless, wearable oximeter: a comparison with arterial SaO2 in healthy volunteers. BMC Anesthesiol. 2022 Mar 24; 22(1): 77.
[9] Schutte AE, Kollias A, Stergiou GS. Blood pressure and its variability: classic and novel measurement techniques. Nat Rev Cardiol. 2022 Oct; 19(10): 643-654.
[10] Schutte AE. Wearable cuffless blood pressure tracking: when will they be good enough? J Hum Hypertens. 2024 Sep; 38(9): 669-672.
[11] Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 327(8476): 307‑310.
[12] Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989; 45(1): 255‑268.
[13] Gaur P, Temple DS, Hegarty-Craver M, Boyce MD, Holt JR, Wenger MF, Preble EA, Eckhoff RP, McCombs MS, Davis-Wilson HC, Walls HJ, Dausch DE. Continuous Monitoring of Heart Rate Variability in Free-Living Conditions Using Wearable Sensors: Exploratory Observational Study. JMIR Form Res. 2024 Aug 7; 8: e53977.
[14] Davies HJ, Williams I, Peters NS, Mandic DP. In-Ear SpO2: A Tool for Wearable, Unobtrusive Monitoring of Core Blood Oxygen Saturation. Sensors (Basel). 2020 Aug 28; 20(17): 4879.
[15] Becking‑Verhaar FL, Verweij RPH, de Vries M, et al. Continuous vital signs monitoring with a wireless device on a general ward: a survey to explore nurses' experiences in a post‑implementation period. Int J Environ Res Public Health. 2023; 20(10): 5794.
[16] Tajrian Amin, Ralph J. Mobbs, Niyaz Mostafa, Luke Sy, Wen Jie Choy. Wearable devices for patient monitoring in the early postoperative period: a literature review mHealth. 2021 Jun 16; 7 (0): 50-50.
[17] Shankar Lal, S Memon, Ehtesham Khan. Adopting the New Norms of Monitoring Postoperative Patients with Wearable Devices in Ireland - Advantages and Challenges Journal of the College of Physicians and Surgeons--Pakistan: JCPSP. 2024 Feb 1; 34 (02): 235-237.
[18] Seongwook Min, Jaehun An, Jae Hee Lee, Ji Hoon Kim, Daniel J. Joe, Soo Hwan Eom, Chang D. Yoo, Hyo–Suk Ahn, Jin-Young Hwang, Sheng Xu, John A. Rogers, Keon Jae Lee. Wearable blood pressure sensors for cardiovascular monitoring and machine learning algorithms for blood pressure estimation Nature reviews. Cardiology. 2025 Feb 18; 0 (0): 0-0.
[19] Aletta E. Schutte. The promise and pitfalls of novel cuffless blood pressure devices European heart journal. 2022 Aug 24; 43 (41): 4222-4223.
[20] Mattia Lunardi, Farooq Muhammad, Atif Shahzad, Asma Nadeem, Lisa Combe, Andrew J. Simpkin, Faisal Sharif, William Wijns, John W. McEvoy. Performance of wearable watch-type home blood pressure measurement devices in a real-world clinical sample Clinical reearch in cardiology: official journal of the German Cardiac Society. 2023 Dec 19; 0 (0): 0-0.
[21] Sheikh Mohammed Shariful Islam, Clara K Chow, Reza Daryabeygi‐Khotbehsara, Narayan Subedi, Jonathan C Rawstorn, Teketo Kassaw Tegegne, Chandan Karmakar, Muhammad Umer Siddiqui, Gavin Lambert, Ralph Maddison. Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis European heart journal. Digital health. 2022 May 2; 3 (2): 323-337.
[22] Yinbo Liu, Carmen C. Y. Poon, Yuan‐Ting Zhang. A hydrostatic calibration method for the design of wearable PAT-based blood pressure monitoring devices Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2008 Aug 1; 0 (0): 1308-1310.
Cite This Article
  • APA Style

    Wang, M., Zhang, P., Ma, W., Hu-Yan, J., Zhang, E. (2026). Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy. Clinical Medicine Research, 15(4), 60-68. https://doi.org/10.11648/j.cmr.20261504.12

    Copy | Download

    ACS Style

    Wang, M.; Zhang, P.; Ma, W.; Hu-Yan, J.; Zhang, E. Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy. Clin. Med. Res. 2026, 15(4), 60-68. doi: 10.11648/j.cmr.20261504.12

    Copy | Download

    AMA Style

    Wang M, Zhang P, Ma W, Hu-Yan J, Zhang E. Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy. Clin Med Res. 2026;15(4):60-68. doi: 10.11648/j.cmr.20261504.12

    Copy | Download

  • @article{10.11648/j.cmr.20261504.12,
      author = {Min Wang and Pan Zhang and Wenhui Ma and Jing Hu-Yan and Erfei Zhang},
      title = {Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy},
      journal = {Clinical Medicine Research},
      volume = {15},
      number = {4},
      pages = {60-68},
      doi = {10.11648/j.cmr.20261504.12},
      url = {https://doi.org/10.11648/j.cmr.20261504.12},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.cmr.20261504.12},
      abstract = {Postoperative vital sign monitoring is essential for early detection of inpatient hemodynamic and respiratory complications. Although conventional bedside monitors serve as the clinical gold standard, their fixed design limits patient mobility and continuous ambulatory monitoring, which hinders early rehabilitation after orthopedic surgery. Wearable monitoring devices have shown great potential for postoperative vital sign surveillance due to their mobility and continuous data collection capabilities, yet real-world validation data in orthopedic postoperative patients remain insufficient, and wearable blood pressure measurement accuracy remains controversial. This study aimed to evaluate the agreement between a wearable monitor and a standard bedside monitor for heart rate, respiratory rate, systolic blood pressure, diastolic blood pressure, and peripheral oxygen saturation (SpO2) in patients undergoing knee arthroscopy. A total of 100 elective knee arthroscopy patients were prospectively enrolled. Paired vital sign data were collected simultaneously via the two devices at four daily time points over two postoperative days. Intra-class correlation coefficient (ICC), Lin’s concordance correlation coefficient, Bland-Altman analysis, temporal trend analysis, and clinical acceptability evaluation were performed for statistical assessment. After excluding five missing values, 3995 paired observations were analyzed. The wearable device showed excellent agreement for heart rate (ICC = 0.994, bias = −0.63 bpm, 95% LoA: −2.97 to 1.70 bpm), good agreement for respiratory rate (ICC = 0.940, bias = −0.51 breaths/min, 95% LoA: −2.71 to 1.68 breaths/min), and moderate-to-good agreement for SpO2 (ICC = 0.872, bias = −0.85%, 95% LoA: −3.50 to 1.80%). Diastolic blood pressure showed moderate agreement (ICC = 0.827, bias = −1.42 mmHg, 95% LoA: −12.90 to 10.06 mmHg), while systolic blood pressure demonstrated the lowest agreement (ICC = 0.811) with a prominent positive systematic bias of 5.63 mmHg. Temporal trends remained stable across postoperative days, with consistent inter-device bias over time. The clinical acceptability rates of heart rate, respiratory rate and SpO2 exceeded 99.5%, compared with 92.2% for diastolic blood pressure and only 79.6% for systolic blood pressure. In conclusion, the wearable monitor provides reliable heart rate, respiratory rate, and SpO2 measurements and is suitable for routine postoperative monitoring after knee arthroscopy. However, blood pressure measurements, particularly systolic blood pressure, show notable systematic deviation, and targeted algorithm calibration is needed to improve its blood pressure monitoring performance.},
     year = {2026}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Agreement Between Wearable-based Vital-sign Monitoring and Standard Bedside Monitoring in Patients After Knee Arthroscopy
    AU  - Min Wang
    AU  - Pan Zhang
    AU  - Wenhui Ma
    AU  - Jing Hu-Yan
    AU  - Erfei Zhang
    Y1  - 2026/10/28
    PY  - 2026
    N1  - https://doi.org/10.11648/j.cmr.20261504.12
    DO  - 10.11648/j.cmr.20261504.12
    T2  - Clinical Medicine Research
    JF  - Clinical Medicine Research
    JO  - Clinical Medicine Research
    SP  - 60
    EP  - 68
    PB  - Science Publishing Group
    SN  - 2326-9057
    UR  - https://doi.org/10.11648/j.cmr.20261504.12
    AB  - Postoperative vital sign monitoring is essential for early detection of inpatient hemodynamic and respiratory complications. Although conventional bedside monitors serve as the clinical gold standard, their fixed design limits patient mobility and continuous ambulatory monitoring, which hinders early rehabilitation after orthopedic surgery. Wearable monitoring devices have shown great potential for postoperative vital sign surveillance due to their mobility and continuous data collection capabilities, yet real-world validation data in orthopedic postoperative patients remain insufficient, and wearable blood pressure measurement accuracy remains controversial. This study aimed to evaluate the agreement between a wearable monitor and a standard bedside monitor for heart rate, respiratory rate, systolic blood pressure, diastolic blood pressure, and peripheral oxygen saturation (SpO2) in patients undergoing knee arthroscopy. A total of 100 elective knee arthroscopy patients were prospectively enrolled. Paired vital sign data were collected simultaneously via the two devices at four daily time points over two postoperative days. Intra-class correlation coefficient (ICC), Lin’s concordance correlation coefficient, Bland-Altman analysis, temporal trend analysis, and clinical acceptability evaluation were performed for statistical assessment. After excluding five missing values, 3995 paired observations were analyzed. The wearable device showed excellent agreement for heart rate (ICC = 0.994, bias = −0.63 bpm, 95% LoA: −2.97 to 1.70 bpm), good agreement for respiratory rate (ICC = 0.940, bias = −0.51 breaths/min, 95% LoA: −2.71 to 1.68 breaths/min), and moderate-to-good agreement for SpO2 (ICC = 0.872, bias = −0.85%, 95% LoA: −3.50 to 1.80%). Diastolic blood pressure showed moderate agreement (ICC = 0.827, bias = −1.42 mmHg, 95% LoA: −12.90 to 10.06 mmHg), while systolic blood pressure demonstrated the lowest agreement (ICC = 0.811) with a prominent positive systematic bias of 5.63 mmHg. Temporal trends remained stable across postoperative days, with consistent inter-device bias over time. The clinical acceptability rates of heart rate, respiratory rate and SpO2 exceeded 99.5%, compared with 92.2% for diastolic blood pressure and only 79.6% for systolic blood pressure. In conclusion, the wearable monitor provides reliable heart rate, respiratory rate, and SpO2 measurements and is suitable for routine postoperative monitoring after knee arthroscopy. However, blood pressure measurements, particularly systolic blood pressure, show notable systematic deviation, and targeted algorithm calibration is needed to improve its blood pressure monitoring performance.
    VL  - 15
    IS  - 4
    ER  - 

    Copy | Download

Author Information
  • Department of Anesthesiology, Yan'an University Affiliated Hospital, Yan’an, China

  • Department of Anesthesiology, Yan'an University Affiliated Hospital, Yan’an, China

  • Department of Anesthesiology, Yan'an University Affiliated Hospital, Yan’an, China

  • Department of Anesthesiology, Yan'an University Affiliated Hospital, Yan’an, China

  • Department of Anesthesiology, Yan'an University Affiliated Hospital, Yan’an, China

  • Sections